Abstract

The ICP-AES technique was developed for rhenium determination in the range of 0.00002 - 0.05 % in copper and molybdenum ores along with their concentrates. The sample pretreatment method by sintering with magnesium oxide and oxidizing agents was offered for analysis; optimum conditions for samples digestion were established. These conditions enabled to pass rhenium into solution with no losses. Matrix influence on rhenium emission spectral lines was studied and methods were investigated to eliminate matrix effects. The analysis results of rhenium-containing ores and concentrates certificated reference materials and mixtures on their basis, obtained by the developed procedure, agreed well with the certified rhenium content values.